The DLAB new generation fully autoclavable manual pipette has the best ergonomic design and is ultralight weight. It has been designed keeping in mind the human form and human factors to offer smooth and effortless ergonomic pipetting experience. It is UV serializable and resistant to strong chemical corrosion. In addition, the volume lock helps protect from volume changing accidentally, and achieve reliable pipetting of smallest volume of liquids.
Single-channel Adjustable Volume Pipettes | |||||||
Cat. No. | Volume Range | Increment | Test Volume | Accuracy error | Precision error | ||
% | μl | % | μl | ||||
7030511005 | 2-20μl | 0.02μl | 20 | 1.00% | 0.2 | 0.30% | 0.06 |
10 | 1.20% | 0.12 | 0.60% | 0.06 | |||
2 | 5.00% | 0.1 | 1.50% | 0.03 | |||
7030511008 | 10-100μl | 0.1μl | 100 | 0.80% | 0.8 | 0.20% | 0.2 |
50 | 1.00% | 0.5 | 0.30% | 0.15 | |||
10 | 3.00% | 0.3 | 1.00% | 0.1 | |||
7030511009 | 20-200μl | 0.2μl | 200 | 0.60% | 1.2 | 0.20% | 0.4 |
100 | 1.00% | 1 | 0.30% | 0.3 | |||
20 | 2.50% | 0.5 | 0.70% | 0.14 | |||
7030511014 | 100-1000μl | 1μl | 1000 | 0.60% | 6 | 0.20% | 2 |
500 | 1.00% | 5 | 0.20% | 1 | |||
100 | 3.00% | 3 | 0.60% | 0.6 |