Feature:
| Specifications | SINGLE-CHANNELADJUSTABLEVOLUMEPIPETTES (10MODELS,COVERING0.1μL~10mL.) |
||||||
| VolumeRange | Increment | TestVolume | SystematicError | RandomError | |||
| μL | μL | μL | μL | % | μL | % | |
| 0.1-2.5 | 0.002 | 2.5 | ±0.04 | ±1.4 | ±0.02 | ±0.7 | |
| 1.25 | ±0.03 | ±2.5 | ±0.02 | ±1.5 | |||
| 0.25 | ±0.03 | ±12.0 | ±0.02 | ±6.0 | |||
| ■0.5-10 | 0.01 | 10 | ±0.10 | ±1.0 | ±0.04 | ±0.4 | |
| 5 | ±0.08 | ±1.5 | ±0.04 | ±0.8 | |||
| 1 | ±0.03 | ±2.5 | ±0.02 | ±1.8 | |||
| ■2-20 | 0.02 | 20 | ±0.20 | ±1.0 | ±0.06 | ±0.3 | |
| 10 | ±0.12 | ±1.2 | ±0.06 | ±0.6 | |||
| 2 | ±0.10 | ±5.0 | ±0.03 | ±1.5 | |||
| ■ 5-50 | 0.05 | 50 | ±0.45 | ±0.9 | ±0.15 | ±0.3 | |
| 25 | ±0.35 | ±1.4 | ±0.20 | ±0.8 | |||
| 5 | ±0.15 | ±3.0 | ±0.08 | ±1.6 | |||
| ■ 10-100 | 0.1 | 100 | ±0.80 | ±0.8 | ±0.20 | ±0.2 | |
| 50 | ±0.50 | ±1.0 | ±0.15 | ±0.3 | |||
| 10 | ±0.30 | ±3.0 | ±0.10 | ±1.0 | |||
| ■ 20-200 | 0.2 | 200 | ±1.20 | ±0.6 | ±0.40 | ±0.2 | |
| 100 | ±1.00 | ±1.0 | ±0.30 | ±0.3 | |||
| 20 | ±0.50 | ±2.5 | ±0.14 | ±0.7 | |||
| ■ 30-300 | 0.2 | 300 | ±1.80 | ±0.6 | ±0.60 | ±0.2 | |
| 150 | ±1.50 | ±1.0 | ±0.45 | ±0.3 | |||
| 30 | ±0.75 | ±2.5 | ±0.21 | ±0.7 | |||
| ■100-1000 | 1 | 1000 | ±6.00 | ±0.6 | ±2.00 | ±0.2 | |
| 500 | ±5.00 | ±1.0 | ±1.00 | ±0.2 | |||
| 100 | ±3.00 |
±3.0 | ±0.60 |
±0.6 |
|||
| ■ 500-5000 | 5 | 5000 |
±30.00 |
±0.6 | ±10.00 |
±0.2 |
|
2500 |
±15.00 |
±0.6 |
±7.50 |
±0.3 |
|||
| 500 | ±12.00 | ±2.4 | ±3.00 | ±0.6 | |||
| ■1000-10000 | 10 | 10000 | ±60.00 | ±0.6 | ±20.00 | ±0.2 | |
| 5000 | ±40.00 | ±0.8 | ±10.00 | ±0.2 | |||
1000 |
±30.00 |
±3.0 |
±6.00 |
±0.6 |
|||
| 7030700000Set 1:0.5-10μL+10-100μL+100-1000μL+Linear Stand+Cap opener+Hipette pen | |||||||
| 7030800000Set 2:2-20μL+20-200μL+100-1000μL+Linear Stand+Cap opener+Hipette pen | |||||||
| 7030900000Set 3:100-1000μL+0.5-5mL+1-10mL+Linear Stand+Cap opener+Hipette pen | |||||||
| User calibration should refer to the industrial standard ISO8655-2. | |||||||